1630 Disposable Medical Pressure Sensor

Brand: American Sensor Technologies Inc

1630 Disposable Medical Pressure Sensor

Brand: American Sensor Technologies Inc
The 1630 is a fully piezoresistive silicon pressure sensor for use in intrauterine pressure monitoring. 

The sensor is designed to be used with automated assembly equipment and can be dropped directly into a customer's disposable intrauterine housing. The sensor is designed to meet the requirements as described in the Association for the Advancement of Medical Instrumentation (AAMI) specification for Blood Pressure Transducers. 

The pressure sensor consists of a pressure sensing element mounted on a ceramic substrate. Thick film resistors on the ceramic substrate are laser trimmed for compensation and calibration. A plastic cap is attached to the ceramic substrate to provide an easy method of attachment to the customer’s assembly and protection for the sensing element. 

A dielectric gel is placed over the sensor to provide fluid isolation. The 1630 pressure sensors are batch manufactured in a 3x8 element array on a ceramic substrate (24 units per substrate). The products are shipped in anti-static shipping containers. 

They can also be shipped on a tape and reel. Performance characteristics and packaging can be easily tailored on a special order basis to meet the requirements of specific customers.
Features
  • Low Cost, Medical Applications
  • Small Size and Reliable Performance
  • Gel Isolation for Liquids
  • Operates from 10 to 40°C
  • Compatible for Automated Assembly
  • 1% Accuracy for Replacements
  • 5.0 uV/V/mmHg Sensitivity
  • Customization for OEM Applications

Specifications

Accuracy

  • 1%

Applications

  • Infusion Pumps
  • Intensive Care Units
  • Intrauterine Monitoring
  • Kidney Dialysis Machines
  • Surgical Procedures
  • Vacuum Assisted Birth

Approvals & Certifications

  • RoHS

Battery Life

  • 3-Years - Shelf Life

Frequency

  • 1,200 Hz

Humidity

  • 10 to 90 %, Non-Condensing - External

Hysteresis

  • ± 2 % of Span

Input Impedance

  • 1,200 to 3,200 ohms

Linearization

  • ± 2 % of Span

Media

  • Dielectric Gel - Media Interface

Operating Temperature

  • 10° to 40° C (50° to 104° F)

Power Requirement

  • 2 UA at 120 Vac RMS, 60 Hz - Leakage Current

Pressure Range

  • mmHg: -50 to 300 (mbar: 67 to 400)

Resistance

  • 270 to 330 ohms - Output

Sensitivity

  • 4.9-5.1 UV/V/mmHg
  • 5 UV/V/mmHg - Typical

Storage Temperature

  • -25° to 70° C (-13° to 158° F)

Supply Voltage

  • 2-10 Vdc / Vac RMS

Thermal Drift

  • ± 0.1 %/°C - Thermal Span Shift
  • ± 0.3 mmHg/°C - Thermal Offset Shift

Update Time

  • 5 seconds - Warm-Up

show more