1620 Disposable Medical Pressure Sensor

Brand: American Sensor Technologies Inc

1620 Disposable Medical Pressure Sensor

Brand: American Sensor Technologies Inc
The 1620 is a fully piezoresistive silicon pressure sensor for use in invasive blood pressure monitoring. 

The sensor is designed to be used with automated assembly equipment and can be dropped directly into a customer's disposable blood pressure housing. The sensor is designed to meet the requirements as described in the Association for the Advancement of Medical Instrumentation (AAMI) specification for Blood Pressure Transducers. 

The pressure sensor consists of a pressure sensing element mounted on a ceramic substrate. Thick film resistors on the ceramic substrate are laser trimmed for compensation and calibration. A plastic cap is attached to the ceramic substrate to provide an easy method of attachment to the customer’s assembly and protection for the sensing element. 

A dielectric gel is placed over the sensor to provide electrical and fluid isolation. The 1620 pressure sensors are batch manufactured in a 10x12 element array on a ceramic substrate (120 units per substrate). The products are shipped in anti-static shipping containers. They can also be shipped on a tape and reel. Performance characteristics and packaging can be easily tailored on a special order basis to meet the requirements of specific customers.
Features
  • Low Cost, Medical Applications
  • Small Size and Reliable Performance
  • Gel Isolation for Liquids
  • Operates from 5 to 40°C
  • Compatible for Automated Assembly
  • 1% Accuracy for Replacements
  • 5.0 uV/V/mmHg Sensitivity
  • Customization for OEM Applications

Specifications

Accuracy

  • 1 % for Replacements

Ambient Temperature Range

  • 23° C (73.4° F)

Applications

  • Infusion Pumps
  • Intensive Care Units
  • Intrauterine Monitoring
  • Kidney Dialysis Machines
  • Surgical Procedures
  • Vacuum Assisted Birth

Approvals & Certifications

  • RoHS

Battery Life

  • 5-Years - Shelf Life

Calibration

  • 102.5 mmHg

Frequency

  • 1,200 Hz

Humidity

  • 10 to 90 % (Non-Condensing)

Hysteresis

  • -1 to 1 mmHg at -30 to 100 mmHg
  • -1.5 to 1.5 % Output at 100 to 300 mmHg

Linearization

  • -1 to 1 mmHg at -30 to 100 mmHg
  • -1.5 to 1.5 % Output at 100 to 300 mmHg

Operating Temperature

  • 5° to 40° C (41° to 104° F)

Operating Time

  • 168 Hours

Power Requirement

  • 10,000 Vdc - Dielectric Breakdown
  • 2 UA at 120 Vac RMS, 60 Hz - Risk Current

Pressure Range

  • mmHg: -50 to 300 (mbar: 67 to 400)

Resistance

  • 1200 to 3200 ohms - Input
  • 285 to 315 ohms - Output

Sensitivity

  • 4.9-5.1 UV/V/mmHg
  • 5 UV/V/mmHg - Typical

Storage Temperature

  • -25° to 70° C (-13° to 158° F)

Supply Voltage

  • 2-10 Vdc / Vac RMS

Thermal Drift

  • ± 0.1 %/°C - Thermal Span Shift
  • ± 0.3 mmHg/°C - Thermal Offset Shift

Update Time

  • 5 seconds - Warm-Up

show more