Approvals & Certifications
Current
- RTD Excitation Current 250 µA
Humidity
- 5 to 95 % - Non-Condensing
Input
- 0-10 Vdc
- 0-20 mA
- 100 mV
- 4-20 mA
- Safe Input Voltage Range (10 Vdc): -0.5 to 15 V
- Safe Input Voltage Range (20 mA): -0.5 to 6 V
- Safe Input Voltage Range (RTD/T/C): ± 24 Vdc
Input Impedance
- 10 kiloohms
- Current Mode: 100 ohms, 35 mA - Max.
- Voltage Mode: 500 kiloohms, 35 mA - Max.
Lower Threshold
- 0.3 mA
- -2.1 mA
- Minimum 3 Vdc
Operating Temperature
- -10° to 60° C (14° to 140° F)
- -22 Heater Option: -40° to 60° C (-40° to 140° F)
Power Requirement
- Inrush 30 A for 1 millisecond at 24 Vdc
- Steady State 130 mA at 24 Vdc
Sensor, Probe Type
- B/R/S Thermocouple: 1600° to 0° C (2912° to 32° F)
- E Thermocouple: 900° to -200° C (1652° to -328° F)
- J Thermocouple: 750° to -210° C (1382° to -346.0° F)
- K/N Thermocouple: 1370° to -240° C (2498° to -400° F)
- T Thermocouple: 400° to -240° C (752° to -400° F)
Storage Temperature
- -10° to 60° C (14° to 140° F)
Supply Voltage
- Maximum
- Minimum Output Supply Voltage: 10 Vdc
Voltage
- Thermocouple Common Mode Range: ± 10 V